Breath Alcohol Analyzer Mistakes Methanol Poisoning for Alcohol Intoxication
January 4, 2010, 11:05 pm





BREATH ALCOHOL ANALYZER MISTAKES METHANOL POISONING FOR ALCOHOL INTOXICATION Caravati EM et al. Ann Emerg Med 2009 Oct 13 [epub ahead of print]
This very interesting case report describes a false positive breath ethanol test erroneously indicating 0.288 g/210 L breath ethanol. In the hospital blood ethanol was not detected; however, the serum methanol concentration was 589 mg/dl. The breath test was done on an Intoxilyzer® 5000EN (CMI Inc). This is a rarely observed false positive result obtained by a breath alcohol analyzer.
The theory behind the false positive is that if a volatile compound has enough vapor pressure to pass from the blood into the breath, and absorbs infrared light at the same wavelength regions as the ethanol molecule, a false reading may result. This is a well written report with a good discussion of the breath analyzer.
Leave a Comment: